pH Test Method |Purpose| Equipment| Test Specimen.



PURPOSE
The main principle is electrometric measurement of the pH value of the aqueous extract of textiles at room temperature by means of a glass electrode. The method is applicable to textiles in any form provided that a representative sample may be obtained which is in a form that permits a ready exchange of liquid between the interior of the material and the solution used in preparing the extract. The result shows the acidity or alkalinity of the textile.

EQUIPMENT
• Distilled water, of at least grade 3, pH 5.0-7.5, conductivity 5 μS/cm
• Standard reference materials buffer (SRM) solutions: pH 6 and 10 or Commercial grade buffer solution: pH 4, pH 7 and pH 10.
• 3 M Potassium Chloride (KCl) solution
• Glass or polypropylene flasks with stopper, 150 ml. The flasks used for this test should be set aside for this purpose only.
• Funnel
• Volumetric flask with stopper, 1000 ml
• Measuring cylinder
• Gloves
• Mechanical shaker, providing rotational or reciprocating movement. A to-and-fro movement at a rate of 60 per minute or a rotational frequency of 30 per minute have been found satisfactory.
• pH-meter, with a glass electrode, capable of measuring to at least 0,1 pH-units
• Balance, accuracy 0.01 g

REAGENTS
• Distilled or deionised water, of at least grade 3, having a pH value between 5.0 and 7.5 and a maximum conductivity of 5 μS/cm.
• Buffer solutions: Either SRM with pH values 6 and 10 or Commercial grade buffer: pH 4, pH 7 and pH 10.
• Potassium chloride solution (KCl) in concentration 0.1 M (mol/litre) prepared using 3 M KCl and distilled water. Below table shows how to prepare 1, 2, 3, 4 or 5 litre of 0.1 M KCl:. To make 1 litre 0.1 M KCl: Take 33 ml of 3 M KCl solution to a 1 litre volumetric flask. Add distilled water up to 1 litre and shake the solution.

Take this volume of 3 M KCl
Take this volume of distilled water
It will give this volume 0.1 M KCl
33 ml
967 ml
1000 ml
67 ml
1933 ml
2000 ml
100 ml
2900 ml
3000 ml
133 ml
3867 ml
4000 ml
167 ml
4833 ml
5000 ml

TEST SPECIMEN
• Each colour on the fabric test specimen shall be tested separately.
• If there are prints, the fabric and print shall be tested together and the colours shall not be separated.

Each test specimen is 2.0 g (±0.05 g).
METHOD
Preparation of samples with mechanical shaker
a) Prepare triplicate samples. Cut (2±0.05)g of the sample that is representative for the bulk production into small pieces having approximately 5 mm sides or of such a size and place them in the flasks. Use gloves.
b) Add 100 ml of 0.1 M KCl solution to each flask.
c) Shake the flasks until the fabric is wet, and then shake it mechanically for 2 hours ± 5 min according to section 5.2.

Preparation of samples without mechanical shaker
a) Prepare triplicate samples. Cut (2±0.05)g of the sample that is representative for the bulk production in small pieces and place them in the flasks. Use gloves.
b) Add 100 ml of 0.1 M KCl solution to each flask.
c) Shake the flasks until the fabric is wet.
d) Let the samples stand for about 15 hours in room temperature. Shake the flasks from time to time.


Calibration of pH meter- with standard reference materials buffer solution
1. Open the fill hole on the electrode; add KCl solutions (3M) if necessary.
2. Switch on the pH meter by pressing [On/Off] key.
3. Rinse the electrode with distilled water and immerse it in the pH 6 buffer solution.
4. Press the [pH cal] key (twice) to measure the temperature.
5. When the display shows “pH 6.00”, press [Enter] to measure the first buffer solution.
6. When the display shows “bu 2”, rinse the electrode, immerse it in the pH 10 buffer solution and press [Enter] to measure the temperature.
7. When the display shows “pH 10.00”, press [Enter] to measure the second buffer solution.
8. When the display shows a slope, the calibration is completed. Note the calibration % on the report sheet. If it is below 93%, calibrate once more with fresh buffer solutions.

Calibration of pH meter - with commercial grade buffer solution (three calibration points)
1. Open the fill hole on the electrode; add KCl solutions (3M) if necessary.
2. Switch on the pH meter by pressing [On/Off] key.
3. Rinse the electrode with distilled water before individual buffer reading.
4. Calibrate with buffer pH 4, pH 7 and pH 10 as per respective pH meter manual.
5. Measure and record the temperature during calibration.

6. When the display shows a slope, the calibration is completed. Note the calibration % on the report sheet. If it is below 93%, calibrate once more with fresh buffer solutions.
Measuring the pH value of the samples
1. Rinse the electrode in distilled water.
2. Place it in flask 1:1 and stir during measuring. Wait for the pH value to stabilize and note the result on the report sheet.
3. Place the electrode in flask 1:2 without rinsing and leave it still. Wait for the triangle in the display to disappear and note the result.
4. Place the electrode in flask 1:3 without rinsing and leave it still. Wait for the triangle in the display to disappear and note the result. The difference between pH values of samples 1:2 and 1:3 should not be greater than 0.3 pH units, if it is the samples need to be measured again.

5. Repeat point 1-4 for all samples.
6. Finally the buffer solutions shall be measured. Rinse the electrode between the two solutions and note the values. If the buffer values differ more than ± 0.05, the calibration and measurement of the samples needs to be re-made.

RESULTS
The average value of second and third measures of one sample is calculated according to the formula below: 𝐴𝑣𝑒𝑟𝑎𝑔𝑒=1:2+1:32
The result shall be reported with one decimal only.

NOTE
Note! If there should be a garment wash on the order, the pH value should be measured after the final treatment is made.
Note! The buffer solutions shall have the same temperature as the samples (room temperature
Note! Sample 1:1 shall not be included in the calculation of the result.


Share :

Facebook Twitter Google+
Back To Top